| # | Tile | Primary Israeli Law / Regulation | Reference No. | Key Authority |
|---|---|---|---|---|
| 1 | Classification Rules | Medical Devices Regulations, Regulation 3 + Schedule 1 | 5770-2010 | MoH / AMAR |
| 2 | Obligations | Medical Devices Regulations, Regulations 4–25 | 5770-2010 | MoH / AMAR |
| 3 | Notified Body | Medical Devices Regulations, Regulation 7(b) + Reliance Pathway | 5770-2010 | MoH / AMAR (Designating Authority) |
| 4 | Battery Requirements | EEE & Batteries Environmental Treatment Law | 5772-2012 | Ministry of Environmental Protection |
| 5 | Shelf-Life Requirements | Medical Devices Regulations, Regulation 9(a)(8) + Schedule 2 | 5770-2010 | MoH / AMAR |
| 6 | REACH | Hazardous Substances Law + Chemical Control Regulations | 5753-1993 | MoH / Ministry of Environmental Protection |
| 7 | RoHS | Restriction of Hazardous Substances in EEE Regulations | 5776-2016 | Ministry of Environmental Protection / SII |
| 8 | Change Control Requirements | Medical Devices Regulations, Regulation 17 | 5770-2010 | MoH / AMAR |
| 9 | OTC Requirements | Medical Devices Regulations, Regulation 9; Consumer Protection Law | 5770-2010 / 5741-1981 | MoH / Consumer Protection & Fair Trade Authority |
| 10 | Export Certificates | Medical Devices Regulations, Regulation 26; MoH Certificate for Export | 5770-2010 | MoH / AMAR |
| 11 | Transportation and Storage | Medical Devices Regulations, Regulation 9(a)(10); Israeli Standards (IS) | 5770-2010 + IS 1552 | MoH / Standards Institution of Israel (SII) |
| 12 | Waste Management | EEE & Batteries Environmental Treatment Law; Public Health Regs (Medical Waste) | 5772-2012 / 5765-2005 | Ministry of Environmental Protection / MoH |
| 13 | AI Medical Device | Medical Devices Regulations (SaMD guidance); IMDRF SaMD N10 | 5770-2010 + MoH Circulars | MoH / AMAR |
Medical Devices Regulations (Takkanot Hatzarot Refuiot), 5770-2010, Regulation 1 — Interpretation:
"'Medical device' means any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes and necessary for its proper application, intended by the manufacturer to be used for human beings for the purpose of: (a) diagnosis, prevention, monitoring, treatment or alleviation of disease; (b) diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap; (c) investigation, replacement or modification of the anatomy or of a physiological process; (d) control of conception; and which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means."
Medical Devices Regulations, Regulation 3 — Classification of Medical Devices:
"Medical devices shall be classified into four classes, according to the rules set out in Schedule 1 of these Regulations, taking into account the intended purpose of the device."
| Class | Risk Level | Examples | Conformity Route |
|---|---|---|---|
| Class I | Low risk | Bandages, examination gloves, non-sterile dressings, walking frames | Self-declaration or CE Mark (Class I sterile / measuring function requires Notified Body) |
| Class IIa | Medium-low risk | Hearing aids, surgical gloves, dental fillings, ultrasound equipment | CE Mark from EU Notified Body / FDA 510(k) or equivalent |
| Class IIb | Medium-high risk | Ventilators, infusion pumps, bone cement, condoms | CE Mark from EU Notified Body / FDA 510(k) or equivalent |
| Class III | High risk | Coronary stents, implantable pacemakers, breast implants, heart valves | CE Mark from EU Notified Body (Design Examination) / FDA PMA or equivalent |
Israel adopts classification rules substantially aligned with the GHTF (Global Harmonization Task Force) Study Group 1 document SG1/N77:2012 — Principles of Medical Devices Classification and the EU Medical Devices Directive (MDD 93/42/EEC) framework, as referenced in Schedule 1 of the Medical Devices Regulations 5770-2010. The rules are based on duration of use (transient, short-term, long-term), invasiveness (non-invasive, invasive, surgically invasive), active/non-active status, and special rules for central circulatory system, CNS, and drug-device combination products.
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 3 + Schedule 1 | www.health.gov.il |
| Public Health Ordinance [New Version], 5740-1984 | Primary enabling law | www.health.gov.il |
| GHTF SG1/N77:2012 — Principles of Medical Devices Classification | Guidance document adopted by Israel MoH | www.imdrf.org |
Obligations in Israel's medical device framework are distributed among four categories of economic operators: Manufacturers, Authorized Representatives (Israeli Authorized Representative — IAR), Importers, and Distributors. All obligations derive from the Medical Devices Regulations, 5770-2010 and subsequent amendments.
Medical Devices Regulations, Regulation 4: "A manufacturer whose registered office is outside Israel shall designate an Authorized Representative established in Israel. The Authorized Representative shall be responsible for registering the device with the Ministry of Health and shall be liable jointly with the manufacturer for the safety and performance of the device placed on the Israeli market."
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulations 4–25 | www.health.gov.il |
| Medical Devices Regulations (Amendment), 5782-2022 | Updated IAR and PMS provisions | www.health.gov.il |
| MoH Director General Circular — Post-Market Surveillance | AMAR Circulars | www.health.gov.il |
Israel does not designate its own Notified Bodies. Instead, the Israeli regulatory system operates on a reliance pathway — accepting conformity assessments and certificates issued by recognized international bodies (primarily EU Notified Bodies and equivalent bodies in reference countries). The MoH/AMAR acts as the sole Designating and Competent Authority.
Medical Devices Regulations, Regulation 7(b): "The Director General may accept evidence of conformity with requirements of a state recognized by the Minister, if that state's requirements are equivalent to the requirements of these Regulations. Recognized states include: European Union member states (CE Marking), United States (FDA), Canada (Health Canada), Australia (TGA), and Japan (PMDA)."
| Reference Country / Region | Accepted Certificate | Notes |
|---|---|---|
| European Union | CE Mark under MDD 93/42/EEC or EU MDR 2017/745 | EU Notified Body certificate accepted; transition to EU MDR ongoing |
| United States | FDA 510(k) Clearance / PMA Approval | Accepted for Class II and III devices |
| Canada | Health Canada Medical Device License | Accepted under bilateral recognition |
| Australia | TGA ARTG Inclusion | Accepted under bilateral recognition |
| Japan | PMDA Shonin / Ninsho | Accepted for Class IIb/III equivalent devices |
| Class | Required Conformity Evidence |
|---|---|
| Class I (non-sterile, non-measuring) | Manufacturer self-declaration (Declaration of Conformity); ISO 13485 QMS certificate |
| Class I (sterile or measuring function) | EU Notified Body involvement for sterility/measuring aspects; or equivalent reference country approval |
| Class IIa / IIb | EU Notified Body certificate (MDD/MDR) or equivalent reference country clearance; ISO 13485 QMS certificate |
| Class III | EU Notified Body Design Examination certificate (EC Type Examination or Full Quality Assurance) or FDA PMA; ISO 13485 QMS certificate |
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 7(b) — Reference Country Reliance | www.health.gov.il |
| MoH Director General Circular — Recognized Foreign Regulatory Authorities | AMAR Circular defining accepted reference countries | www.health.gov.il |
| IMDRF GHTF/SG1/N78:2012 — Principles of Conformity Assessment for Medical Devices | Adopted by Israel MoH as reference guidance | www.imdrf.org |
Environmental Treatment of Electrical and Electronic Equipment and Batteries Law (Chok Tipul Svivati b'Tziyud Chashmali v'Electroni u'Bataryot), 5772-2012, Section 1 — Definitions:
"'Battery' means a source of electrical energy generated by direct conversion of chemical energy, consisting of one or more primary battery cells, including any batteries or accumulators placed on the market as part of an electrical or electronic device. 'Battery for medical use' includes any battery incorporated in a medical device or used to power a medical device."
Medical devices powered by or incorporating batteries — including implantable devices with batteries (e.g., pacemakers, neurostimulators), portable battery-powered devices (e.g., portable monitors, infusion pumps), and home-use medical devices — are subject to the battery requirements of the 5772-2012 Law and the associated Regulations issued thereunder by the Ministry of Environmental Protection.
| Document | Reference | Source URL |
|---|---|---|
| Environmental Treatment of Electrical and Electronic Equipment and Batteries Law, 5772-2012 | Primary law governing battery requirements | www.gov.il (Ministry of Environmental Protection) |
| Regulations for Environmental Treatment of Batteries, issued under Law 5772-2012 | Implementing regulations — substance limits, labeling, collection targets | www.gov.il (Ministry of Environmental Protection) |
| Medical Devices Regulations, 5770-2010, Schedule 2 (Essential Requirements) | Section on devices incorporating energy sources — safety requirements for battery-powered devices | www.health.gov.il |
Medical Devices Regulations, 5770-2010, Regulation 9(a)(8) — Labeling Requirements:
"The labeling of a medical device shall include: (8) the expiry date, expressed at minimum as year and month, after which the device is not to be used, or, where applicable, the date of manufacture including the year and month of manufacture. If there is no relevant expiry date, the label shall state the shelf life from the date of manufacture."
Medical Devices Regulations, 5770-2010, Schedule 2 — Essential Requirements, Section 9:
"Devices shall be designed, manufactured and packaged in such a way that their characteristics and performances during their intended use will not be adversely affected during storage and transport, taking account of the instructions and information to be provided by the manufacturer. Manufacturers shall indicate the shelf life applicable to the device, including to sterile devices, and specify the conditions under which the device must be stored to maintain its performance."
Israel adopts ISO standards through the Standards Institution of Israel (Makhon HaTaknim HaIsraeli — SII). Relevant adopted standards include:
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 9(a)(8) — Labeling; Schedule 2, Section 9 — Storage and shelf life | www.health.gov.il |
| IS EN ISO 11607-1:2019 (adopted by SII) | Packaging for terminally sterilized medical devices | www.sii.org.il |
| GHTF/SG1/N70:2011 — Label and Instructions for Use for Medical Devices | Guidance adopted by MoH/AMAR for labeling content | www.imdrf.org |
Israel does not have a direct equivalent to the EU REACH Regulation (EC) No 1907/2006. Chemical management in Israel is governed by a separate legislative framework based on the Hazardous Substances Law, 5753-1993 and associated regulations, administered jointly by the Ministry of Health (MoH) and the Ministry of Environmental Protection. The Standards Institution of Israel (SII) also plays a role in adopting relevant international chemical safety standards.
Hazardous Substances Law (Chok Hahomrim HaMasuchenim), 5753-1993, Section 1 — Purpose:
"The purpose of this Law is to regulate the import, export, manufacture, transport, storage, use, and disposal of hazardous substances, to prevent harm to persons and the environment, and to ensure the safe handling of chemical substances in Israel."
Hazardous Substances Regulations (Labeling and Packaging), 5768-2008:
"Hazardous substances placed on the market in Israel shall be classified, labelled, and packaged in accordance with the provisions of these Regulations, which are based on the GHS (Globally Harmonized System of Classification and Labelling of Chemicals) as adapted for Israel."
Israel's chemical control framework does not replicate EU REACH's full scope (registration, evaluation, authorization). Manufacturers exporting to Israel should note that REACH compliance (for EU market) does not automatically fulfil Israeli requirements — however, substances restricted under EU REACH Annex XVII that overlap with Israeli Hazardous Substances Register restrictions must be separately verified against Israeli law.
| Document | Reference | Source URL |
|---|---|---|
| Hazardous Substances Law, 5753-1993 | Primary chemical control law | www.gov.il (MoH) |
| Hazardous Substances Regulations (Labeling and Packaging), 5768-2008 | GHS-based labeling requirements | www.gov.il (Ministry of Environmental Protection) |
| Medical Devices Regulations, 5770-2010, Schedule 2, Section 7 | Chemical substances in medical devices — essential requirement | www.health.gov.il |
Restriction of Hazardous Substances in Electrical and Electronic Equipment Regulations (Takkanot Hagbalat Shimush b'Homrim Masuchenim b'Tziyud Chashmali v'Electroni), 5776-2016, Regulation 1 — Purpose:
"The purpose of these Regulations is to restrict the use of certain hazardous substances in electrical and electronic equipment placed on the Israeli market, in order to protect human health and the environment, taking into account the environmentally sound recovery and disposal of such equipment."
Restriction of Hazardous Substances Regulations, 5776-2016, Regulation 3 — Restricted Substances:
"Electrical and electronic equipment placed on the Israeli market shall not contain the following substances in excess of the maximum concentration values: (a) Lead (Pb) — 0.1%; (b) Mercury (Hg) — 0.1%; (c) Cadmium (Cd) — 0.01%; (d) Hexavalent Chromium (Cr VI) — 0.1%; (e) Polybrominated Biphenyls (PBB) — 0.1%; (f) Polybrominated Diphenyl Ethers (PBDE) — 0.1%; (g) Bis(2-ethylhexyl) phthalate (DEHP) — 0.1%; (h) Butyl benzyl phthalate (BBP) — 0.1%; (i) Dibutyl phthalate (DBP) — 0.1%; (j) Diisobutyl phthalate (DIBP) — 0.1%."
Israel's RoHS Regulations (5776-2016) are modeled on EU RoHS 2 (2011/65/EU) including the 2015 amendment (Directive 2015/863/EU) which added the four phthalates. The regulations apply to electrical and electronic medical devices — i.e., all active medical devices (Class I active, IIa, IIb, III) that contain electrical or electronic components. Non-electrical/non-active medical devices are outside the scope of RoHS.
| Document | Reference | Source URL |
|---|---|---|
| Restriction of Hazardous Substances in EEE Regulations, 5776-2016 | Primary RoHS legislation (Israel) | www.gov.il (Ministry of Environmental Protection) |
| Environmental Treatment of EEE and Batteries Law, 5772-2012 | Enabling legislation for RoHS Regulations | www.gov.il (Ministry of Environmental Protection) |
| Standards Institution of Israel (SII) — IS EN IEC 63000:2018 | Technical documentation for RoHS assessment of EEE (adopted by SII) | www.sii.org.il |
Medical Devices Regulations, 5770-2010, Regulation 17 — Changes to Registered Devices:
"The holder of a registration certificate for a medical device shall notify the Director General of any change to the device, its manufacturing process, its indications, its labeling, or its Instructions for Use, prior to implementing such change on the Israeli market. The Director General shall determine, in accordance with the nature and significance of the change, whether the change requires a new registration application or may be implemented following notification and approval."
| Change Type | Category | Required Action |
|---|---|---|
| Changes affecting safety, performance, or intended purpose | Significant / Major | New registration application or Variation application — prior approval required from MoH/AMAR before implementation |
| Changes to labeling (content, language), IFU updates, addition/removal of contraindications | Moderate | Notification to MoH/AMAR; await approval letter before implementation |
| Administrative changes (company name, address, contact details of IAR) | Minor / Administrative | Notification to MoH/AMAR; may implement after submission (acknowledgment sufficient) |
| Manufacturing process changes (materials, sterilization parameters) | Significant | Prior approval required; may require updated Notified Body certificate / reference country evidence |
| Software updates to SaMD (Software as a Medical Device) | Risk-dependent | Evaluated per IMDRF SaMD N23 guidance; significant functionality changes require new registration or variation |
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 17 — Changes to Registered Devices | www.health.gov.il |
| MoH Director General Circular — Guidance on Change Notifications for Registered Medical Devices | AMAR Circular providing classification table for change types | www.health.gov.il |
| IMDRF/SaMD WG/N23 FINAL:2015 — Software as a Medical Device: Possible Framework for Risk Categorization and Corresponding Considerations | Guidance adopted for SaMD change assessment | www.imdrf.org |
Over-the-counter (OTC) medical devices in Israel are devices available for purchase by the general public without a prescription or healthcare professional intermediary. These devices are subject to the general medical device registration requirements plus specific labeling, advertising, and consumer protection obligations.
Medical Devices Regulations, 5770-2010, Regulation 9(b) — OTC Labeling:
"For medical devices intended to be used by the lay public (general consumer), the labeling and Instructions for Use shall be written in Hebrew, in clear and understandable language, including: (1) the intended purpose of the device; (2) a description of how to use the device safely; (3) warnings and contraindications; (4) information on when to consult a healthcare professional; (5) expiry date and storage conditions; (6) the name and address of the Israeli Authorized Representative."
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 9(b) — OTC Labeling Requirements | www.health.gov.il |
| Consumer Protection Law (Chok Hagant HaTzarchan), 5741-1981 | General consumer protection obligations for OTC products | www.gov.il (Consumer Protection Authority) |
| MoH Director General Circular — Advertising of Medical Devices to the Public | Specific guidance on permitted OTC advertising claims | www.health.gov.il |
Medical Devices Regulations, 5770-2010, Regulation 26 — Certificate for Export:
"The Director General may, upon application, issue a certificate in respect of a medical device manufactured in Israel for the purpose of export, attesting to the status of the device under Israeli law, the identity of the manufacturer, and whether the device is registered in the Israeli Medical Device Registry. Such certificate may be issued in the format recommended by the World Health Organization or in any other format approved by the importing country's regulatory authority."
| Certificate Type | Issued By | Purpose |
|---|---|---|
| Certificate for Free Sale (CFS) | MoH / AMAR | Confirms device is legally marketed in Israel; required by many import countries for registration |
| Certificate of Conformity (CoC) | MoH / AMAR | Confirms device meets Israeli regulatory requirements and is registered in IMDR |
| WHO-format Certificate | MoH / AMAR | WHO-type certificate for export to countries requiring WHO format (commonly requested by developing markets) |
| GMP Certificate (ISO 13485) | Accredited Certification Body (not MoH) | ISO 13485 QMS certificate — issued by accredited third-party certification bodies, not MoH |
(For further details on fees, please refer to the Timeline and Fees section.)
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 26 — Certificate for Export | www.health.gov.il |
| AMAR Online Portal — Export Certificate Application | Online submission portal for export certificates | www.health.gov.il |
| WHO Guidelines for Inspection of Drug Manufacturers and for Import/Export of Pharmaceutical Products | WHO certificate format adopted by MoH | www.who.int |
Medical Devices Regulations, 5770-2010, Schedule 2 (Essential Requirements), Section 9 — Storage and Transportation:
"Devices shall be designed, manufactured and packaged in such a way that their characteristics and performances during their intended use will not be adversely affected during transport and storage, taking account of the instructions and information provided by the manufacturer. The manufacturer shall specify the conditions of storage and transport necessary to ensure proper performance of the device throughout its intended lifetime."
Medical Devices Regulations, 5770-2010, Regulation 9(a)(10) — Labeling:
"The label of a medical device shall indicate: (10) special conditions for storage and/or handling, including the temperature range, humidity conditions, pressure conditions, or any other environmental conditions necessary to maintain the integrity and performance of the device."
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Schedule 2, Section 9 (Storage/Transport ESR); Regulation 9(a)(10) (Labeling) | www.health.gov.il |
| IS EN ISO 11607-1:2019 (SII) | Sterile barrier system packaging validation | www.sii.org.il |
| Israeli Dangerous Goods Transport Regulations (implementing ADR) | Transport of devices containing hazardous materials, batteries | www.gov.il (Ministry of Transport) |
Medical device waste management in Israel is governed by two parallel frameworks: (1) the medical waste regulations administered by the MoH (for clinical/biological waste generated by use of medical devices in healthcare settings), and (2) the WEEE (Waste Electrical and Electronic Equipment) framework administered by the Ministry of Environmental Protection (for electrical/electronic medical device end-of-life waste).
Environmental Treatment of Electrical and Electronic Equipment and Batteries Law (Chok Tipul Svivati b'Tziyud Chashmali v'Electroni u'Bataryot), 5772-2012, Section 3 — Producer Obligations:
"Producers of electrical and electronic equipment placed on the Israeli market are responsible for financing and organizing the collection and treatment of waste from their equipment. Producers shall register with the Ministry of Environmental Protection and join an approved collective compliance scheme or establish an individual producer responsibility system. Medical equipment falling within the categories defined in Schedule 1 of this Law is subject to the obligations of this Law."
Public Health Regulations (Medical Waste), 5765-2005, Regulation 1 — Definitions:
"'Medical waste' means waste generated in the course of providing medical care or health services, including waste contaminated with blood, body fluids, pathological waste, sharps, and microbiological cultures. Medical waste shall be segregated, packaged, labeled, transported, and disposed of in accordance with these Regulations and the guidelines issued by the Director General of the Ministry of Health."
| Category | Applicable Law | Responsible Party | Key Obligation |
|---|---|---|---|
| Electrical/electronic medical device WEEE | EEE & Batteries Law 5772-2012 | Manufacturer / Importer (Producer) | Register with MoEnP; join collective compliance scheme; provide free take-back at end of life; achieve collection/recycling targets |
| Clinical/biological medical waste | Public Health Regs (Medical Waste) 5765-2005 | Healthcare facilities / waste generators | Segregate at source; use approved yellow/red biohazard containers; contract licensed medical waste treatment company; maintain disposal records |
| Sharps waste (needles, lancets from devices) | Public Health Regs (Medical Waste) 5765-2005 | Healthcare facility / household user (community programs) | Dispose in approved puncture-resistant sharps containers; community sharps disposal programs available for home users |
| Radioactive device waste (e.g., brachytherapy seeds) | Radiation Protection Law 5719-1959 + Medical Devices Regs | Manufacturer / Authorized User | Disposal under Israel Atomic Energy Commission (IAEC) regulations |
| Document | Reference | Source URL |
|---|---|---|
| Environmental Treatment of EEE and Batteries Law, 5772-2012 | Sections 3–12 — WEEE producer obligations | www.gov.il (Ministry of Environmental Protection) |
| Public Health Regulations (Medical Waste), 5765-2005 | Full regulation — clinical waste segregation, treatment, disposal | www.health.gov.il |
| EEE and Batteries Environmental Treatment Regulations (Implementation), 5774-2014 | Implementing regulations setting collection targets and reporting obligations | www.gov.il (Ministry of Environmental Protection) |
Israel does not yet have dedicated AI-specific medical device legislation. AI-enabled medical devices (including machine learning-based Software as a Medical Device — SaMD) are regulated as medical devices under the existing Medical Devices Regulations, 5770-2010, with classification and conformity requirements applied based on the intended purpose and risk class. The MoH/AMAR follows international frameworks from IMDRF and adopts guidance documents relevant to SaMD and AI/ML.
Medical Devices Regulations, 5770-2010, Regulation 1 — Definition (Software as a Medical Device):
"The definition of 'medical device' in these Regulations includes software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes, when such software independently performs one or more medical purposes as defined herein, without being part of a hardware medical device. AI-based software that meets this criterion is regulated as a medical device."
MoH Director General Circular — Digital Health Technology and Software as a Medical Device (Chozer Mankal L'Technologiat Briut Digitalit):
"Software as a Medical Device (SaMD), including software based on artificial intelligence and machine learning algorithms, is subject to registration requirements under the Medical Devices Regulations, 5770-2010. Classification of SaMD follows the IMDRF SaMD N12 framework. Changes to AI/ML algorithms that affect the intended purpose or performance of the device require prior notification or re-registration."
Israel adopts the IMDRF SaMD classification framework:
| IMDRF SaMD Category | Equivalent Israeli MD Class | Healthcare Situation | Examples |
|---|---|---|---|
| Category IV (highest risk) | Class III | Critical / treat or diagnose | AI-based clinical decision support for life-threatening conditions; autonomous diagnostic AI (e.g., AI ECG interpretation affecting treatment in ICU) |
| Category III | Class IIb | Serious / treat or diagnose | AI-based radiology image analysis for cancer detection; AI-based drug dosing recommendation systems |
| Category II | Class IIa | Non-serious / treat or diagnose; or Serious / inform | AI-based screening tools for non-critical conditions; AI-based triage tools providing information to clinicians |
| Category I (lowest risk) | Class I | Non-serious / inform | Administrative software, patient wellness apps not meeting SaMD definition; general health tracking software |
Israel's MoH/AMAR is monitoring the EU AI Act (Regulation (EU) 2024/1689) and FDA's AI/ML-based SaMD Action Plan. Dedicated AI medical device guidance or legislation is anticipated. Manufacturers should monitor MoH Director General Circulars (Chozrei Mankal) for updates. Israel's active involvement in IMDRF working groups suggests alignment with IMDRF N97 (AI/ML-enabled Medical Devices) framework is likely in upcoming guidance.
| Document | Reference | Source URL |
|---|---|---|
| Medical Devices Regulations, 5770-2010 | Regulation 1 (SaMD definition); Schedule 2 (Essential Requirements for software) | www.health.gov.il |
| IMDRF/SaMD WG/N12 FINAL:2014 — Software as a Medical Device: Possible Framework for Risk Categorization | SaMD classification framework adopted by Israel MoH | www.imdrf.org |
| IMDRF/SaMD WG/N23 FINAL:2015 — Software as a Medical Device: Possible Framework for Risk Categorization and Corresponding Considerations | Guidance on SaMD lifecycle changes including AI algorithm updates | www.imdrf.org |
| IMDRF/CYBER WG/N60 FINAL:2020 — Principles and Practices for Medical Device Cybersecurity | Cybersecurity requirements for connected SaMD/AI devices | www.imdrf.org |
| MoH Director General Circular — Digital Health Technology and SaMD | Israeli-specific guidance on SaMD registration and AI classification | www.health.gov.il |