Austria — SaMD Full Gap Analysis

RegDesk RegIntel · Competent Authority: BASG (Bundesamt für Sicherheit im Gesundheitswesen) · Product: Software as a Medical Device (SaMD)

Generated: September 9, 2026  |  Framework: EU MDR 2017/745 + MPG 2021 + EU AI Act 2024/1689

Austria — SaMD Regulatory Gap Analysis

Competent Authority: BASG — Bundesamt für Sicherheit im Gesundheitswesen
Secondary Authority: AGES MEA (Austrian Medicines & Medical Devices Agency)
National Registry: Medizinprodukteregister (Austrian Medical Device Registry)
Governing Law: EU MDR 2017/745 · Medizinproduktegesetz 2021 (MPG 2021) · Medical Devices Reporting Ordinance 2024
SaMD Classification: Typically Class IIa–III under EU MDR Rule 11 (software used for diagnosis/therapy decisions)
AI-enabled SaMD: Also subject to EU AI Act 2024/1689 (high-risk AI system from August 2026)


📌 SaMD Context Note: Austria follows EU MDR directly — there is no separate national SaMD approval pathway. CE marking + EUDAMED registration is the primary market access mechanism. BASG acts as market surveillance authority. The EU AI Act adds a dual-compliance layer for AI-enabled SaMD from August 2026.

Gap Analysis Summary

21
Tiles Assessed
11
✅ Covered
6
⚠️ Partial / Needs Update
4
❌ Gap / Missing
StatusCountTile Sections
✅ Covered11Classification, CE Marking, QMS, Clinical Evaluation, Labeling (General), UDI, PMS/PMCF, Vigilance/AE Reporting, Authorized Representative, Importation, EUDAMED
⚠️ Partial6National Registration, Fees, Clinical Investigations, SaMD-Specific Requirements, Language Requirements, Distributor Obligations
❌ Gap4EU AI Act / AI-enabled SaMD, eSubmission / EUDAMED Modules (mandatory 2026), Renewal / Certificate Lifecycle, Custom-Made Device Rules
Total21

1. Regulatory Framework & Classification

✅ Covered

What's in Place

SaMD Classification Detail

SaMD UseClassConformity Assessment
Driving clinical decisions, diagnosis, treatment planningIIa / IIb / IIINotified Body required
Monitoring / alerting with no direct therapy impactIIaNotified Body required
General wellness / administrative (no diagnostic claim)ISelf-declaration

Source: BASG Medical Devices · EU MDR 2017/745 · MDX CRO SaMD Guide


2. CE Marking & Conformity Assessment

✅ Covered

Source: BASG · OMC Medical


3. National Registration

⚠️ Partial — Needs Update (rules changed May 2024)

Current State (post May 1, 2024)

The Medical Devices Reporting Ordinance 2024 (effective May 1, 2024) significantly changed national registration obligations:

Gap / Action Needed

Backend tile should reflect the post-May 2024 reform: old content stating all actors must register nationally is now outdated. EUDAMED compliance replaces national registration for most SaMD manufacturers.

Source: Medizinprodukteregister FAQs · BASG · Preslmayr Law 2024


4. Quality Management System (QMS)

✅ Covered

Source: SimplerQMS · Maven Prof Serv


5. Clinical Evaluation

✅ Covered

Source: BASG · EU MDR 2017/745 Annex XIV


6. Clinical Investigations (Clinical Trials)

⚠️ Partial — Incomplete Detail on SaMD-Specific Pathway

What Applies

Gap

Backend tile lacks SaMD-specific context: for SaMD, "clinical investigation" may mean usability studies, retrospective data collection, or algorithm validation studies — these nuances (MDCG 2020-1 guidance on clinical evaluation for SaMD) should be in the tile.

Source: BASG Clinical Investigations


7. Labeling Requirements

⚠️ Partial — Language Requirement Detail Missing

General Labeling

Austria-Specific Language Requirements

Gap

Language requirement granularity (consumer vs. professional) should be explicitly documented per the BASG position.

Source: BASG Importer · Arazy Group Austria Guide


8. Unique Device Identification (UDI)

✅ Covered

Source: BASG · MDI Europa


9. Authorized Representative

✅ Covered

Source: BASG · EU MDR Art. 11


10. Importation

✅ Covered

Source: BASG Importer page · EU MDR Art. 13–14


11. Distributor Obligations

⚠️ Partial — National Registration Requirement Needs Clarification

Gap

Backend tile should clarify that while manufacturer registration shifted to EUDAMED, distributors in Austria still must register nationally. This distinction is a common source of confusion post-2024.

Source: Medizinprodukteregister FAQs · EU MDR Art. 14


12. Post-Market Surveillance (PMS) & PMCF

✅ Covered

Source: BASG Market Surveillance · EU MDR Art. 83–86


13. Adverse Events / Vigilance Reporting

✅ Covered

Source: BASG Reporting · EU MDR Art. 87–92


14. EUDAMED (European Database for Medical Devices)

✅ Covered (but rapidly evolving — needs monitoring)

Source: BASG · European Commission EUDAMED


15. Fees & Annual Declarations

⚠️ Partial — Fee Schedule Detail Needed

Annual Fee Structure

Device ClassAnnual Fee (approx.)Notes
Class ILower tierSelf-declaration sufficient
Class IIaMid tier—
Class IIb~€350/yearHighest class reported fee
Class IIIHighest tier—

Gap

Full fee schedule (all classes + 2025 tariffs) should be extracted from BASG fee page and added to backend tile with exact amounts.

Source: BASG Medical Devices Fee · Preslmayr Law 2024


16. Renewal / Certificate Lifecycle

❌ Gap — Not in Backend

What Applies

Gap

No tile exists for certificate lifecycle management, NB certificate renewal timelines, or the MDR significant change assessment process for SaMD. This is particularly important for SaMD given frequent software updates.

Source: BASG · MDCG 2020-3 · EU MDR Art. 120


17. SaMD-Specific Requirements

⚠️ Partial — Lacks SaMD-Specific Tile

Key SaMD Requirements Under EU MDR (Austria)

Gap

Backend currently lacks a dedicated SaMD tile. Recommend creating a "Software as a Medical Device" section under Austria with the above requirements and MDCG guideline references.

Source: MDX CRO SaMD Guide · Maven Prof Serv · MDCG 2019-16 · MDCG 2020-3


18. EU AI Act — AI-Enabled SaMD

❌ Gap — Critical Missing Tile

Regulatory Requirements

Key Timelines

DateObligation
August 2025General-purpose AI (GPAI) model obligations apply (relevant if SaMD built on foundation model)
August 2026High-risk AI obligations fully apply — new AI-powered SaMD must comply
2027+Full enforcement across all AI Act provisions

High-Risk AI Requirements (for AI-enabled SaMD)

Gap

No tile exists in Austria backend for EU AI Act obligations on SaMD. This is a critical gap given the August 2026 deadline is imminent.

Source: MDX CRO EU AI Act · EU AI Act 2024/1689 · MDCG 2025-6


19. eSubmission / EUDAMED Module Compliance

❌ Gap — Evolving Requirements Not Captured

Gap

Backend tile should document EUDAMED module rollout schedule, Austria-specific implications, and the transition from national to EU-level registration.

Source: BASG · European Commission EUDAMED roadmap


20. Custom-Made Device Rules

❌ Gap — Specific Tile Missing

Gap

No tile distinguishing custom-made device obligations from standard MDR pathway — particularly relevant given custom-made SaMD use cases in Austria's healthcare system.

Source: Medizinprodukteregister · EU MDR Annex XIII


21. Timeline & Key Dates Summary

✅ Covered (partial — should be expanded)

DateRequirementImpact on SaMD
May 27, 2024MDD/AIMDD certificates expiredAll SaMD must be MDR-certified
May 1, 2024Medical Devices Reporting Ordinance 2024National registration rules changed
May 26, 2025UDI mandatory for Class I SaMDSoftware UDI implementation required
August 2025EU AI Act GPAI obligationsFoundation-model-based SaMD affected
May 2026EUDAMED mandatory for PMS/VigilanceReporting workflow change required
May 28, 2026EUDAMED UDI registration mandatory (new devices)Registration system change
August 2026EU AI Act high-risk AI obligationsAI-SaMD must comply (critical deadline)
November 27, 2026EUDAMED UDI registration mandatory (existing devices)Legacy device update deadline
June 30 (annual)BASG fee declaration deadlineAll SaMD suppliers in Austria

Priority Action Items for Backend Team

PriorityTile / AreaAction Required
P0 — CriticalEU AI Act / AI-enabled SaMDCreate new tile — August 2026 deadline imminent; all AI-SaMD affected
P0 — CriticalEUDAMED Module RolloutUpdate tile with 2025–2026 mandatory deadlines and Austria transition rules
P1 — HighNational Registration (post-May 2024)Update to reflect reform: manufacturers → EUDAMED; distributors → still national
P1 — HighSaMD-Specific RequirementsCreate dedicated SaMD tile (IEC 62304, 62366, cybersecurity, versioning, eIFU)
P1 — HighCertificate Lifecycle / RenewalAdd tile: NB certificate renewal, significant change assessment for SaMD
P2 — MediumFeesAdd exact fee schedule per class from BASG 2025 tariff; add exemption rules
P2 — MediumLanguage RequirementsClarify consumer vs. professional-use distinction for German language requirement
P2 — MediumDistributor ObligationsClarify that distributors still require national registration (unlike manufacturers)
P2 — MediumCustom-Made DevicesAdd tile: Austrian custom-made device rules, national registration, SaMD use case
P3 — LowClinical InvestigationsAdd SaMD-specific guidance (usability studies, retrospective data, MDCG 2020-1)

Key Sources & References

SourceURLType
BASG — Main Medical Devices Portalbasg.gv.at/en/medical-devicesOfficial / Competent Authority
BASG — Importer Informationbasg.gv.at/en/medical-devices/importerOfficial
BASG — Medical Devices Feebasg.gv.at/en/medical-devices/medical-devices-feeOfficial
BASG — Clinical Investigationsbasg.gv.at clinical-trials-with-medical-devicesOfficial
BASG — Market Surveillance & Reportingbasg.gv.at/en/market-surveillance/reportingOfficial
Medizinprodukteregister (Austrian MD Registry)medizinprodukteregister.at/FAQsOfficial National Registry
EU MDR 2017/745eur-lex.europa.euEU Regulation
EU AI Act 2024/1689eur-lex.europa.euEU Regulation
Preslmayr Law — Austria Pharma & MD Regulation 2024preslmayr.at (PDF)Legal Commentary
CMS Law — Austria What's New in Medical Device Lawcms.lawLegal Update
MDCG 2025-6 — MDAI Guidanceec.europa.euEU Guidance
Arazy Group — Austria IVD/MD Registration Guidearazygroup.comIndustry Guide
OMC Medical — Austria Registrationomcmedical.comIndustry Guide